凯时K66

//jujiaopai.com/en/index.php/product/index.html">

A PHP Error was encountered

Severity: Notice

Message: Undefined index: title

Filename: inc/header.php

Line Number: 108

  • Glove Integrity Tester
  • Sterility Test
  • //jujiaopai.com/en/index.php/product/index.html">

    A PHP Error was encountered

    Severity: Notice

    Message: Undefined index: title

    Filename: inc/header.php

    Line Number: 108

  • Sample Pre-treatment
  • Microbial Limit Test
  • Endoscope Microbial Test
  • TOC Analyzer
  • Filter Integrity Tester
  • Media Degasser
  • Services
  • News
  • Innovations
  • About TAILIN
  • //jujiaopai.com/en/index.php/product/index.html">

    A PHP Error was encountered

    Severity: Notice

    Message: Undefined index: title

    Filename: inc/header.php

    Line Number: 194

  • Glove Integrity Tester
  • Sterility Test
  • Sample Pre-treatment
  • Microbial Limit Test
  • Endoscope Microbial Test
  • TOC Analyzer
  • Filter Integrity Tester
  • Media Degasser
  • Services
  • News
  • Innovations
  • About TAILIN
  • CN
  • Popular Searches
    Rigid Sterility Test Isolator
    Rigid Sterility Test Isolator
    Rigid Sterility Test Isolator
    Rigid Sterility Test Isolator
    Rigid Sterility Test Isolator
    Rigid Sterility Test Isolator
    Rigid Sterility Test Isolator
    Rigid Sterility Test Isolator
    Rigid Sterility Test Isolator


    The laminar flow aseptic isolator composed of stainless steel and toughened glass.  It is Class A inside.

    The system is equipped with Siemens (PLC) for full automatic control.

    Comly with GMP, EP, USP,CP

    According to GMP, the sterility test is allowed to be made in the sterility test isolator which is installed in the Class D background.


    Features

    1.   Built-in Vaporized Hydrogen Peroxide (VHPS) generator for bio-decontamination which can reach 6-log Sterility Assurance Level;

    2.   The VHPS residual concentration is less than 1ppm after aeration;

    3.   Leakage rate is less than 0.5% /vol/hour under 2 times working pressure test;

    4.   Pressure is adjustable within -80pa~80pa;

    5.   12.1” IPC with NetSCADA;

    6.   Built-in Sterility Test Pump;

    7.   Built-in Glove Integrity Tester (HTY-GIT01INS);

    8.   Comply with 21 CFR Part 11 electronic records and signature requirements;

    9.   Interlock, alarms record and audit trails;

    10. Noise: ≤65dB(A) when in pressure hold stage

    11. Compressed air supply: 0.4MPa~0.6MPa clean and dry

    12. Power Supply: 380±38V,50±5Hz,25A ( HTY-1800G4/HTY-1800AG4)


    Options

    Air Sampler, Particle Counter

    High / low concentration sensor

    Portable concentration sensor for operator safe

    WiFi controlled glove integrity tester (GIT-WLAN)

    Wooden Mock-Up



    Applications

    Aseptic filling, sampling,sterility testing, compounding and dispensing.

    Specifications

    ModelDescription

    Overall Dimensions (L*W*H)

    Operating SpaceInner Pass Box
    HTY-1250G2

    Main Chamber + Pass Box

    - Single Side

    2300*870*2420 mm1240*680*700 mm600*550*700 mm
    HTY-1250AG2Main Chamber - Single Side1670*870*2420 mm1240*680*700 mm          /
    HTY-1650G3

    Main Chamber + Pass Box

    - Single Side

    2700*870*2420 mm1640*680*700 mm600*550*700 mm
    HTY-1650AG3Main Chamber - Single Side2075*870*2420 mm1640*680*700 mm        /
    HTY-1800G4

    Main Chamber + Pass Box

    - Single Side

    2850*870*2420 mm1790*680*700 mm600*550*700 mm
    HTY-1800AG4Main Chamber - Single Side2225*870*2420 mm1790*680*700 mm       /
    HTY-1800G8

    Main Chamber + Pass Box

    - Double Side


    2650*1240*2420 mm1750*1200*650 mm600*550*700 mm
    HTY-1800AG8Main Chamber - Double Side2370*1240*2420 mm1750*1200*650 mm       /

    * Size can be customized

    Services

    Complete validation protocols to comply with FDA, EMA, GMP, 21CFR Part 11.

    FAT, SAT, Training.

     

     